  • Title: Calibration Method for a High-Density3D Photoacoustic Tomography Imaging System in the Research of Breast Cancer

Abstract

Breast cancer stands as the most prevalent cancer and the second leading cause of cancer-related deaths among women. Approximately 20% of breast cancer cases fall into the triple-negative breast cancer type (TNBC), typically identified in later stages and associated with higher chances of local and distant recurrence, leading to a shorter overall survival time. Given these findings from clinical studies, there exists an urgent demand for an efficient surgical procedure to ensure complete removal of tumor lesions, thus reducing the likelihood of recurrence both locally and distantly. Traditional breast imaging methods such as X-ray, ultrasound, and magnetic resonance imaging (MRI) are limited. Consequently, there is a crucial need for innovative imaging approaches to aid in the accurate and sensitive detection of tumor lesions for surgical removal. Photoacoustic Tomography (PAT), as one of the rapidly advancing biomedical optical imaging techniques, offers the advantage of nonionizing and non-invasive imaging of tissue absorption, delivering high spatial resolution. This technology has already demonstrated its potential in both biological and clinical applications, making it a promising candidate for addressing the challenges in breast cancer detection and treatment.

In this research, we introduce an innovative real-time functional 3D Photoacoustic Tomography (PAT) imaging system equipped with 1,024 convex virtual point transducers. To ensure accurate positioning of each transducer element, we propose a new calibration method utilizing a standard micro-ball with a diameter of 0.2mm. Briefly, the calibration procedure involves the following steps:  Step 1: An ideal point target represented by a micro-ball is imaged, and the center coordinate (center of gravity) of the micro-ball is calculated based on the reconstructed PAT images without any prior calibration. Step 2: Calibration data, representing the displacement of each transducer element, is generated by comparing the measured delay time of the signal with the calculated delay time based on the target position obtained from Step 1. Step 3: The calibration process is repeated multiple times, with the target positioned at different locations. By averaging the results from all these experiments, we generate the final calibration data for the entire system. To validate the efficacy of this new calibration method, we conducted several experiments using static phantoms (employing pencil leads and human hairs as targets) and dynamic phantoms (using a micro-tube with ink solution flow).

Furthermore, we imaged two tumor-bearing mice, systematically injected with tumor-targeted NIR fluorescent probes (NIR830-ATF-IONP), to further demonstrate the effectiveness of this calibration technique and showcase the imaging capabilities of our 3D PAT system post-calibration. The results indicate that this calibration method significantly enhances image quality and can be widely applied to calibrating PAT systems with numerous transducer elements. Moreover, these experiments highlight the tremendous potential of our novel 3D PAT system as a powerful imaging tool with high spatial and temporal resolution, particularly in the study of breast cancer.